• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN SURGICAL, S.A. OPTILENE 6/0 (0,7) 75CM 2XDR13B CV2 RCP; CARDIOVASCULAR SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN SURGICAL, S.A. OPTILENE 6/0 (0,7) 75CM 2XDR13B CV2 RCP; CARDIOVASCULAR SUTURE Back to Search Results
Model Number C3097191
Device Problems Material Frayed (1262); Material Integrity Problem (2978); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2023
Event Type  malfunction  
Manufacturer Narrative
Reported device not marketed in the u.S., however, similar devices or devices that share components, raw materials, process methods or other technological characteristics are registered within the u.S.Under pma/510(k) number: k133890 if additional information becomes available a follow up report will be submitted.
 
Event Description
It was reported an issue with optilene suture.The client reported that during a right carotid endarterioectomy (arteriography) the polypropylene thread was frayed and subsequently ruptured.There was no patient injury nor consequences reported.
 
Manufacturer Narrative
Analysis and results: there are no previous complaints of this code-batch.We manufactured and distributed in the market (b)(4) units of this code-batch.There are no units in stock in b.Braun surgical's warehouse.We have received one open and used sample in a pot with rests of blood and with the thread damaged.In these conditions, this sample cannot be checked.Furthermore, without any closed sample we cannot carry out an analysis in order to take a decision.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfil usp/ep and b.Braun surgical requirements.We have also reviewed the complaint history record, and there are two previous complaints of other references and batches with the same thread raw material batch as the used in this product regarding another issue in the last five years, closed as not confirmed after the analysis.As indicated in the instructions for use of the product, when working with optilene suture material, great care should be taken to avoid any crushing or crimping damage of the monofilament by instruments such as forceps or needle holders.Final conclusion: in spite of receiving a defective sample, without closed samples a suitable analysis cannot be performed.Nevertheless, we take note of this incidence and if any closed sample is received in the future, we will re-open the case and analyse it.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTILENE 6/0 (0,7) 75CM 2XDR13B CV2 RCP
Type of Device
CARDIOVASCULAR SUTURE
Manufacturer (Section D)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP  08191
Manufacturer (Section G)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
Manufacturer Contact
martina laporte
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
MDR Report Key16494201
MDR Text Key310805730
Report Number3003639970-2023-00045
Device Sequence Number1
Product Code GAW
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
SEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC3097191
Device Catalogue NumberC3097191
Device Lot Number121504
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/23/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
-
-