Reported device not marketed in the u.S., however, similar devices or devices that share components, raw materials, process methods or other technological characteristics are registered within the u.S.Under pma/510(k) number: k133890 if additional information becomes available a follow up report will be submitted.
|
Analysis and results: there are no previous complaints of this code-batch.We manufactured and distributed in the market (b)(4) units of this code-batch.There are no units in stock in b.Braun surgical's warehouse.We have received one open and used sample in a pot with rests of blood and with the thread damaged.In these conditions, this sample cannot be checked.Furthermore, without any closed sample we cannot carry out an analysis in order to take a decision.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfil usp/ep and b.Braun surgical requirements.We have also reviewed the complaint history record, and there are two previous complaints of other references and batches with the same thread raw material batch as the used in this product regarding another issue in the last five years, closed as not confirmed after the analysis.As indicated in the instructions for use of the product, when working with optilene suture material, great care should be taken to avoid any crushing or crimping damage of the monofilament by instruments such as forceps or needle holders.Final conclusion: in spite of receiving a defective sample, without closed samples a suitable analysis cannot be performed.Nevertheless, we take note of this incidence and if any closed sample is received in the future, we will re-open the case and analyse it.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
|