Model Number D134804
Device Problems Tear, Rip or Hole in Device Packaging (2385); Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2023
Event Type  malfunction  
Manufacturer Narrative
The product has not returned for analysis, however, a picture(s) were provided by the customer.Evaluation is still in progress.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Event Description
It was reported that a patient was to undergo a cardiac ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter however, the device presented with damaged in the packaging and the sterility was compromised.It was reported that a catheter was delivered to the account with damaged packaging causing compromised sterility.There was both damage to the shipment box and the outside.It does appear that the device secured properly in the tray.Also noted there was a dent on the catheter.The hard plastic side of the catheter packing that was damaged/perforated.There were no patient consequences; the device was not used on a patient.
 
Manufacturer Narrative
On 21-mar-2023, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Manufacturer Narrative
It was reported that a patient was to undergo a cardiac ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter however, the device presented with damaged in the packaging and the sterility was compromised.It was reported that a catheter was delivered to the account with damaged packaging causing compromised sterility.There was both damage to the shipment box and the outside.It does appear that the device secured properly in the tray.Also noted there was a dent on the catheter.The hard plastic side of the catheter packing that was damaged/perforated.There were no patient consequences; the device was not used on a patient.Device evaluation details: the device was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.A visual inspection of the returned device was performed following bwi procedures.Visual analysis revealed damage in the packaging, it was observed broken; which could compromise the sterility of the catheter.The damage could be related to the process of storing, packaging or shipping of the device, however, this cannot be conclusively determined.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key16499689
MDR Text Key311378942
Report Number2029046-2023-00488
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 05/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD134804
Device Catalogue NumberD134804
Device Lot Number30928457L
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received04/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-