• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH BATTERY HANDPIECE/MODULAR FOR TRS; BATTERY, REPLACEMENT, RECHARGEABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH BATTERY HANDPIECE/MODULAR FOR TRS; BATTERY, REPLACEMENT, RECHARGEABLE Back to Search Results
Catalog Number 05.001.201
Device Problems Mechanical Jam (2983); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese, or its employees that the report constitutes an admission that the product, depuy synthese, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition that the device had moving parts that did not move smoothly-trigger, identified during service and repair, was confirmed.The assignable root cause was determined to be due to component failure from wear.Udi: (b)(4).
 
Event Description
It was reported by japan that during service and evaluation, it was determined that the battery handpiece device had a leak tightness test failure, moving parts did not move smoothly-trigger, would not reverse, would not hold/secure/battery, deformed/bent housing and component damage.It was further determined that the device failed pretest for general condition, leakage test using bubble emission technique, check for sticky triggers, check roundness of housing, check falling out protection (steel ring) and check general function of device.It was noted that the device drop ring was not in the correct position.It was noted in the service order that the device did not reverse.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY HANDPIECE/MODULAR FOR TRS
Type of Device
BATTERY, REPLACEMENT, RECHARGEABLE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key16500938
MDR Text Key310949454
Report Number8030965-2023-02833
Device Sequence Number1
Product Code MOQ
UDI-Device Identifier07611819977822
UDI-Public07611819977822
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2023
Date Manufacturer Received03/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/14/2011
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-