Model Number 384466
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2023
Event Type  malfunction  
Event Description
¿date of picc insertion: (b)(6) 2022¿.¿date of event: (b)(6) 2023¿.¿description of event: rn noted leaking from the blue lumen near the tubing/hub connection¿.¿a new picc was placed form completion of prescribed therapy¿.
 
Manufacturer Narrative
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
 
Event Description
Date of picc insertion ¿ (b)(6) 2022¿.¿date of event ¿ 1/20/2023¿.¿description of event ¿ rn noted leaking from the blue lumen near the tubing/hub connection.¿ ¿a new picc was placed form completion of prescribed therapy.¿.
 
Manufacturer Narrative
A review of the dhr and inspection records was conducted, and no similar concerns were found.One opened catheter was returned for review.Visual inspection confirmed a crack in the luer that would result in leakage.Several complaints for this part number have previously been received regarding a cracked hub resulting in leakage, and capa 2021-039 was initiated to address this issue.The capa is currently in the implementation phase and will evaluate the corrective action implementation for effectiveness to prevent a recurrence of this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L-CATH PICC D/L 20GA (2.6F) 0.90MM X 60CM
Type of Device
L-CATH PICC
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
elnaz rahman
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key16501635
MDR Text Key310951276
Report Number0001625425-2023-00991
Device Sequence Number1
Product Code LJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091670
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number384466
Device Catalogue Number384466
Device Lot Number11431205
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/24/2023
Initial Date FDA Received03/08/2023
Supplement Dates Manufacturer Received02/27/2023
Supplement Dates FDA Received04/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-