• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. TURBO-CLEANING CLOSED SUCTION SYSTEM FOR ADULTS, 14 F, DSE; VAP CLOSED SUCTION CATHETERS & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. TURBO-CLEANING CLOSED SUCTION SYSTEM FOR ADULTS, 14 F, DSE; VAP CLOSED SUCTION CATHETERS & ACCESSORIES Back to Search Results
Model Number 227-5
Device Problem Ejection Problem (4009)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2023
Event Type  malfunction  
Manufacturer Narrative
A review of the device history record is in-progress.The product involved in the report has been returned and the investigation remains in progress at this time.All information reasonably known as of 07 mar 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to (b)(4), in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Event Description
It was reported, the locking system on the thumb valve got stuck (down) in the suction mode and the button had to be pulled out manually.It wouldn't spring out as it should when locking.No injury or medical interventions reported.
 
Manufacturer Narrative
The device history record for lot 30105098 was reviewed and the product was produced according to product specifications.The actual complaint product was returned for evaluation.The entire device was visually inspected for damage, no breakages or damages were seen at the outer portion of the device.The position of the thumb valve was in the upright position when received.The thumb valve was pressed and released multiple times for functional inspection.Each time after release, the thumb valve returned to the upright position.The thumb valve was also inspected if the rotation was possible, multiple times (3 times), the thumb valve was rotated from the unlocked to locked position, and there were no difficulties or issues seen.The samples were also measured for dimensions on the control valve body (cvb) and seal components.An interference was observed between the seal and cvb.The incident was confirmed as reported.A root cause could not be determined.All information reasonably known as of 17 may 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURBO-CLEANING CLOSED SUCTION SYSTEM FOR ADULTS, 14 F, DSE
Type of Device
VAP CLOSED SUCTION CATHETERS & ACCESSORIES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S DE RL DE CV
carretera internacional
salida norte no. 1053
magdalena, cp
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key16506399
MDR Text Key311245529
Report Number8030647-2023-00042
Device Sequence Number1
Product Code BSY
UDI-Device Identifier00609038989655
UDI-Public00609038989655
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number227-5
Device Catalogue NumberN/A
Device Lot Number30105098
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-