Model Number 096F6P |
Device Problems
Difficult or Delayed Positioning (1157); Difficult to Advance (2920)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
As reported, during the last 3 months when performing right catheterizations, in some cases, the 6f swan ganz catheters used did not advance smoothly once they have passed the introducer, leading to difficulties to pass through the central vein and heart cavities.Some occasions the issue was solved by changing to femoral approach, in some other occasions the introducer was replaced with another one from a different brand using the same approach but the issue remained.
|
|
Manufacturer Narrative
|
We were informed that the swan ganz catheter was not received at our product evaluation laboratory for a full evaluation since it was discarded.Without return of the product, edwards is unable to perform a complete investigation of the reported event.It is not possible to determine what factors may have contributed to it, and therefore no actions could be planned.The lot number for this device was not supplied; therefore, further review of the related manufacturing records could not be performed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|
|
Manufacturer Narrative
|
Based on further engineering investigation, the units go through a balloon inflation inspection as part of the manufacturing process.Current risk mitigations include design and manufacturing controls, ifu warnings and cautions, and physician training of device insertion techniques.
|
|
Search Alerts/Recalls
|
|
|