MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ CATHETERS; CATHETER, FLOW DIRECTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ CATHETERS; CATHETER, FLOW DIRECTED Back to Search Results
Model Number 096F6P
Device Problems Difficult or Delayed Positioning (1157); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
As reported, during the last 3 months when performing right catheterizations, in some cases, the 6f swan ganz catheters used did not advance smoothly once they have passed the introducer, leading to difficulties to pass through the central vein and heart cavities.Some occasions the issue was solved by changing to femoral approach, in some other occasions the introducer was replaced with another one from a different brand using the same approach but the issue remained.
 
Manufacturer Narrative
We were informed that the swan ganz catheter was not received at our product evaluation laboratory for a full evaluation since it was discarded.Without return of the product, edwards is unable to perform a complete investigation of the reported event.It is not possible to determine what factors may have contributed to it, and therefore no actions could be planned.The lot number for this device was not supplied; therefore, further review of the related manufacturing records could not be performed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Manufacturer Narrative
Based on further engineering investigation, the units go through a balloon inflation inspection as part of the manufacturing process.Current risk mitigations include design and manufacturing controls, ifu warnings and cautions, and physician training of device insertion techniques.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CATHETERS
Type of Device
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
samantha eveleigh
one edwards way
irvine, CA 92614
9492503939
MDR Report Key16512912
MDR Text Key311361078
Report Number2015691-2023-11347
Device Sequence Number1
Product Code DYG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193466
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number096F6P
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-