Model Number LIFEVEST WCD 4000 SYSTEM |
Device Problems
Charging Problem (2892); Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/13/2023 |
Event Type
malfunction
|
Event Description
|
A us distributor reported that a battery charger was unable to recognize a battery pack.
|
|
Manufacturer Narrative
|
Device evaluation of battery charger has been completed.The reported problem (battery charger problem) was due to corrosion on the battery board.The root cause for the corrosion was ingress of an unknown liquid.Lifevest patient training materials have been updated as a reminder not to expose the lifevest electronic components to liquids.There was no adverse event that resulted from the damaged charger.
|
|