Model Number 3100A VENT,115V,60HZ,DOM |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/14/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
A vyaire field service engineer checked the unit and ran performance check.Found delta p at 82.Calibrated ddi (dynamic displacement indicator) and pr2 (pressure regulator).Delta p within specifications at 67 afterwards.Vyaire medical proceeded with completing 2k pm.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
It was reported to vyaire medical problem with 3100a ventilator where the customer observed that the patient was not improving, so they swapped the unit out with another 3100a ventilator.They just want to make sure that the unit is functioning properly.Furthermore, there was no patient harm reported associated with the reported event.
|
|