An event of device deformity during deployment was reported.A returned device assessment could not be performed as the device was not returned for analysis.Possible causes of device deformity include use of the incorrect size delivery system, anatomical interference, and angulation or kink in the delivery system upon deployment.Information from the field indicated that there were no kinks noticed in the delivery system and there was no interaction with cardiac structures during the deployment.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.The field indicated that the device was implanted in it's elongated state.Please note per the instructions for use, if device position is unsatisfactory: stabilize the wire and re-advance the delivery catheter until the device is completely within the catheter.Reposition and deploy, or remove the device from the patient.
|