• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 150 - DEU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 150 - DEU Back to Search Results
Model Number 27053
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2023
Event Type  malfunction  
Manufacturer Narrative
The astral device was returned to resmed.Evaluation confirmed the reported complaint.Visual inspection of the main circuit board revealed super capacitor leakage.The main circuit board will be replaced to address the reported complaint.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed an error message (sf179) related to the super capacitor.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - DEU
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key16517528
MDR Text Key311038999
Report Number3007573469-2023-00153
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00619498270538
UDI-Public(01)00619498270538(11)160601(10)1179369
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27053
Device Catalogue Number27053
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/09/2023
Distributor Facility Aware Date02/10/2023
Device Age80 MO
Date Report to Manufacturer03/09/2023
Date Manufacturer Received02/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/06/2016
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-