Device Problems Calibration Problem (2890); Problem with Software Installation (3013)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2023
Event Type  malfunction  
Event Description
It was reported that the programmer was unable to update.The programmer was returned for service and subsequently tested out of specification during manufacturers analysis. there was no patient involvement.
 
Manufacturer Narrative
Product analysis: analysis was able to confirm the customer comment that the programmer was unable to update.It was also determined at analysis that the cooling fan was not working, and the company logo was missing.The upper left hinge was worn, and errors were found in the software history.Th radiofrequency head cable was worn but worked correctly.The rf anti-slip layer was missing.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC B.V.
earl bakkenstraat 10
heerlen 6422 PJ
NL  6422 PJ
Manufacturer (Section G)
MEDTRONIC B.V.
earl bakkenstraat 10
heerlen 6422 PJ
NL   6422 PJ
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key16520546
MDR Text Key311049915
Report Number3002807561-2023-00002
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090WDE
Device Catalogue Number2090WDE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2023
Date Manufacturer Received03/03/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-