MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC HEARTSTRING III SYSTEM 3.8MM; CLAMP, VASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC HEARTSTRING III SYSTEM 3.8MM; CLAMP, VASCULAR Back to Search Results
Model Number HSK-3038
Device Problems Premature Activation (1484); Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/13/2023
Event Type  malfunction  
Manufacturer Narrative
Trackwise id (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
 
Event Description
The hospital reported that during a coronary artery bypass procedure, the hst iii system (3.8mm) fell off and remained in the loading device when the delivery system was withdrawn from the loading device.The seal was not loaded into the delivery system.Steps 2 and 3 were good.It is possible that the load in step 4 was insufficient.White plunger pressed.The procedure was completed with other manufacturer's products (used enclosure), and there was no extension of the operation time.The patient had no health hazards.
 
Manufacturer Narrative
Trackwise#: (b)(4).The lot #25161829 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTRING III SYSTEM 3.8MM
Type of Device
CLAMP, VASCULAR
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key16529188
MDR Text Key311580532
Report Number2242352-2023-00182
Device Sequence Number1
Product Code DXC
UDI-Device Identifier00607567700314
UDI-Public00607567700314
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K130382
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/15/2022
Device Model NumberHSK-3038
Device Catalogue NumberHSK-3038
Device Lot Number25161829
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/30/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/15/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ENCLOSURE
-
-