Model Number HSK-3038 |
Device Problems
Premature Activation (1484); Fitting Problem (2183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/13/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Trackwise id (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
|
|
Event Description
|
The hospital reported that during a coronary artery bypass procedure, the hst iii system (3.8mm) fell off and remained in the loading device when the delivery system was withdrawn from the loading device.The seal was not loaded into the delivery system.Steps 2 and 3 were good.It is possible that the load in step 4 was insufficient.White plunger pressed.The procedure was completed with other manufacturer's products (used enclosure), and there was no extension of the operation time.The patient had no health hazards.
|
|
Manufacturer Narrative
|
Trackwise#: (b)(4).The lot #25161829 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|
|