Model Number N/A |
Device Problem
Packaging Problem (3007)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/17/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the distributor found the outer packaging was crushed and there was damage to the sterile package.There was no patient involvement.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
(b)(4).Report source: japan.Product is in process of being returned to zimmer biomet for the investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
No additional event information to report at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Visual evaluation of the returned product found a puncture mark in the paperboard carton.The puncture mark penetrated the carton and created stress marks on both the inner and outer sterile blisters.The stress marks do not compromise the sterile barrier.Sterility has not been breached.This complaint has been confirmed by review of the returned product.Review of the device history record(s) identified no deviations or anomalies during manufacturing.The condition of the device when it left zimmer biomet is conforming to specification.The root cause of the reported event can be attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|