• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Increase in Pressure (1491)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/15/2023
Event Type  malfunction  
Event Description
Philips received a complaint on the v60 ventilator indicating that a 1009 (vent-inop: pressure regulation high) error code occurred during preventative maintenance (pm).The device was reported to be outside of use at the time of the reported problem.No patient harm was reported.The customer was provided with a quote for a replacement gas delivery system (gds).The quote is pending acceptance or rejection by the customer.This investigation is ongoing.
 
Manufacturer Narrative
Philips was unable to confirm the final disposition of the device because the customer rejected the quote, refusing service.No further work was performed by philips to resolve the issue.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key16543499
MDR Text Key311376559
Report Number2518422-2023-06720
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020047
UDI-Public00884838020047
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Device Catalogue Number1053616
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/24/2023
Date Device Manufactured08/17/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-