The customer from (b)(6) obtained a discordant elevated advia centaur cp - total hcg (thcg) result for one female patient sample.The initial result was reported to the physician, who questioned the result.The sample was repeated on the same advia centaur system, and the result was lower.The interpretation of results section of the advia centaur cp total hcg instructions for use states: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." siemens continues to investigate.
|
The customer obtained a discordant elevated advia centaur cp - total hcg (thcg) result for one female patient sample.The initial result was reported to the physician, who questioned the results.The sample was repeated on the same advia centaur system, and the result was lower.There are no reports that treatment was altered or prescribed or adverse health consequences due to the discordant advia centaur cp total hcg result.
|
Initial mdr 1219913-2023-00049 was filed on march 15, 2023 reporting a customer from united states obtained a discordant elevated advia centaur cp - total hcg (thcg) result for one female patient sample.The initial result was reported to the physician, who questioned the result.The sample was repeated on the same advia centaur cp system, and the result was lower.Additional information on march 22, 2023: service replaced the reagent syringe on (b)(6) 2023.The customer had no further concerns with the instrument or thcg performance since service.No further support is required.Additional information on march 24, 2023: the customer reported a falsely elevated discordant patient result on advia centaur cp total hcg (thcg) reagent lot 345.On (b)(6) 2023, a patient resulted as 570 miu/ml which was questioned by the physician.The same sample was repeated and resulted as <5 miu/ml.The lower repeat value was issued with a corrected report.The calibration of thcg reagent lot 345 was valid and biorad quality control (qc) lots 40400 and 42293 recovered within range on the day of this event.The customer pulled thcg samples that ran before and after the discordant sample and all three of those samples repeated similar to the initial result.Service was dispatched and learned that on (b)(6) 2023 the operators cleaned up a substantial amount of a liquid substance from the top of the incubation ring cover.There was evidence of a spill, and it is possible a few of cuvettes were contaminated in the vicinity of the liquid.The service engineer suspected that the liquid could have come from the reagent syringe assembly so it was replaced.Following service thcg was recalibrated, quality control (qc) was in range and no leaks have been identified.The customer is operational and requires no further support.No potential product issue is observed.The customer is operational.
|