|
Lot Number 0030706441 |
Device Problems
Leak/Splash (1354); Off-Label Use (1494)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/16/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the sheath exhibited a hemostatic valve leak.During a procedure a polarsheath was selected for use.All cooling was completed, the polarx cryoablation catheter was removed, and a non-bsc ring catheter was inserted (which constitutes off-label use of the sheath); however, blood slowly started flowing from the hemostatic valve of the sheath.Since there was not enough blood loss to require replacement, the procedure was completed using the device as it was.No patient complications were reported.The device has been returned for analysis.
|
|
Manufacturer Narrative
|
The polarsheath was returned to boston scientific for laboratory analysis.A visual inspection did not reveal any damage that would lead to blood leaking from the hemostatic valve; however, the hemostatic valve was found to have a large tear which was visible under microscopy.This tear is likely the cause for the leak seen in the field.The tear was separate from the manufacturing slit.Functional testing was also performed, which the sheath failed in some parameters.During a leakage test the device was gently pressurized with 6 psi at the flushing line lure fitting while plugging the distal tip of the sheath, and the pressure decay value for the sheath passed.The device passed a hemostatic valve test without a dilator inserted.The sheath was able to maintain a minimum pressure of 5.5 psi.No leaks were observed.During an aspiration test no air was visible in the flushing line during test/syringe vacuum, though the device failed a 10 cc syringe with draw (approx 1 cc per second) test as well as a 60 cc syringe with draw (approx 30 cc per second) test.Bubbles were visible in the flushing line during test/syringe vacuum in those tests.The polarsheath was used with a ring catheter (non-bsc device) used during procedure which constitutes off label use.Laboratory analysis confirmed the reported clinical observation.
|
|
Event Description
|
It was reported that the sheath exhibited a hemostatic valve leak.During a procedure a polarsheath was selected for use.All cooling was completed, the polarx cryoablation catheter was removed, and a non-bsc ring catheter was inserted (which constitutes off-label use of the sheath); however, blood slowly started flowing from the hemostatic valve of the sheath.Since there was not enough blood loss to require replacement, the procedure was completed using the device as it was.No patient complications were reported.The device has been returned for analysis.
|
|
Event Description
|
It was reported that the sheath exhibited a hemostatic valve leak.During a procedure a polarsheath was selected for use.All cooling was completed, the polarx cryoablation catheter was removed, and a non-bsc ring catheter was inserted (which constitutes off-label use of the sheath); however, blood slowly started flowing from the hemostatic valve of the sheath.Since there was not enough blood loss to require replacement, the procedure was completed using the device as it was.No patient complications were reported.The device has been returned for analysis.
|
|
Manufacturer Narrative
|
The polarsheath was returned to boston scientific for laboratory analysis.A visual inspection did not reveal any damage that would lead to blood leaking from the hemostatic valve; however, the hemostatic valve was found to have a large tear which was visible under microscopy.This tear is likely the cause for the leak seen in the field.The tear was separate from the manufacturing slit.Functional testing was also performed, which the sheath failed in some parameters.During a leakage test the device was gently pressurized with 6 psi at the flushing line lure fitting while plugging the distal tip of the sheath, and the pressure decay value for the sheath passed.The device passed a hemostatic valve test without a dilator inserted.The sheath was able to maintain a minimum pressure of 5.5 psi.No leaks were observed.During an aspiration test no air was visible in the flushing line during test/syringe vacuum, though the device failed a 10 cc syringe with draw (approx 1 cc per second) test as well as a 60 cc syringe with draw (approx 30 cc per second) test.Bubbles were visible in the flushing line during test/syringe vacuum in those tests.The polarsheath was used with a ring catheter (non-bsc device) used during procedure which constitutes off label use.Laboratory analysis confirmed the reported clinical observation.
|
|
Search Alerts/Recalls
|
|
|