• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL LIVEWIRE¿ ELECTROPHYSIOLOGY CATHETER DUO-DECAPOLAR (20 ELECTRODES), SUPER LARGE; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL LIVEWIRE¿ ELECTROPHYSIOLOGY CATHETER DUO-DECAPOLAR (20 ELECTRODES), SUPER LARGE; CATHETER, STEERABLE Back to Search Results
Model Number 401904
Device Problem Entrapment of Device (1212)
Patient Problem Foreign Body In Patient (2687)
Event Date 03/03/2023
Event Type  Injury  
Event Description
During insertion of the catheter, it became stuck in the atrium and the pull wires broke.A snare was used to removed the catheter.The catheter was retained by the customer.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported entrapment could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIVEWIRE¿ ELECTROPHYSIOLOGY CATHETER DUO-DECAPOLAR (20 ELECTRODES), SUPER LARGE
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345
Manufacturer (Section G)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key16550882
MDR Text Key311371515
Report Number2182269-2023-00012
Device Sequence Number1
Product Code DRA
UDI-Device Identifier05414734202534
UDI-Public05414734202534
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K022380
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number401904
Device Catalogue Number401904
Device Lot Number8744969
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-