LIVANOVA DEUTSCHLAND S5 ROLLER PUMP PRO; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 10-80-00Z |
Device Problem
Pumping Stopped (1503)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/15/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova deutschland manufactures the s5 roller pump pro.The incident occurred in san antonio, texas.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.In order to fix it, motor control board and front panel of the s5 roller pump pro were replaced.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a s5 roller pump pro gave two motor control failure error messages at setup.There was no patient involvement.
|
|
Manufacturer Narrative
|
H10: the involved pump is installed since 2021 and according to the review of the complaints database, no further event has been reported for this unit.Based on the information available, it cannot be excluded that the root cause could be traced back to an electrical/electronic deviation involving the replaced parts.
|
|
Event Description
|
See initial report.
|
|