MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP PRO; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP PRO; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00Z
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/15/2023
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova deutschland manufactures the s5 roller pump pro.The incident occurred in san antonio, texas.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.In order to fix it, motor control board and front panel of the s5 roller pump pro were replaced.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a s5 roller pump pro gave two motor control failure error messages at setup.There was no patient involvement.
 
Manufacturer Narrative
H10: the involved pump is installed since 2021 and according to the review of the complaints database, no further event has been reported for this unit.Based on the information available, it cannot be excluded that the root cause could be traced back to an electrical/electronic deviation involving the replaced parts.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP PRO
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key16551468
MDR Text Key311657347
Report Number9611109-2023-00108
Device Sequence Number1
Product Code DWB
UDI-Device Identifier04033817902713
UDI-Public(01)04033817902713(11)210930
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00Z
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/30/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-