• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND VELYS SAW INTERFACE LEFT; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND VELYS SAW INTERFACE LEFT; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 4515-70-105
Device Problems Loose or Intermittent Connection (1371); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation of the returned device found that it was loose.Visual examination found that there were no signs of damage or improper usage.The overall appearance had signs of multiple uses.A review of the device history record indicates that there were no non conformances relevant to the current complaint condition.The issue related to the sasi being loose has been escalated to a capa.The assignable root cause could not be determined.Udi: (b)(4).
 
Event Description
It was reported that the robotic assisted saw interface device was "bad or the saw was bad".During an in-house engineering evaluation, it was determined that the device was found to be loose.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in the surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELYS SAW INTERFACE LEFT
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
DEPUY IRELAND
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key16551848
MDR Text Key311384032
Report Number1818910-2023-05896
Device Sequence Number1
Product Code OLO
UDI-Device Identifier10603295519461
UDI-Public10603295519461
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K202769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4515-70-105
Device Catalogue Number451570105
Device Lot NumberJ43865
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2023
Date Manufacturer Received03/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-