Manufacturer's investigation conclusion: the reported event of system controller failing to alarm was not confirmed.The system controller periodic log file spanned approximately 255 days ((b)(6) 2022 to (b)(6) 2023 per the timestamp).Ebb usage count increased from 12 to 69 between (b)(6) 2022 at 14:41 and (b)(6) 2023 at 14:20 due to loss of external power.Per provided information, the usage of backup battery was identified to be caused by faulty outlet while the device was connected to mobile power unit (mpu); however, the patient did not observe associating no external power alarm when the backup battery was used to operate the system.The periodic log file captures events at specific interval and did not capture the onset of the events.No other notable alarm was observed.The hm3 system controller, serial (b)(4), was not returned for analysis.A root cause of the reported event was not determined through this analysis.Device history records were reviewed and showed no deviations from manufacturing or qa specifications.The patient handbook cautions the users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.Heartmate iii instructions for use section 7-¿alarms and troubleshooting¿ and heartmate iii patient handbook section 5-¿alarms and troubleshooting¿ explain how to properly interpret and troubleshoot alarms.No further information was provided.The manufacturer is closing the file on this event.
|