Catalog Number 04625374160 |
Device Problems
Circuit Failure (1089); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/16/2023 |
Event Type
malfunction
|
Event Description
|
We received an allegation of questionable inr results for 1 patient tested with coaguchek xs meter serial number (b)(4) compared to another coaguchek xs meter (clinic's meter) unknown serial number.On (b)(6)2023 at 11:20 a.M.The patient had a clinic's meter result of 3.2 inr while at 11:35 a.M.The patient's meter result was 2.3 inr.The patient's therapeutic range was 2.5-3.5 inr and the interval of testing was every 3 weeks.The lot number and the expiration date of the test strips used on the clinic's meter were not provided.
|
|
Manufacturer Narrative
|
Occupation is patient/consumer.The patient's and the clinic's test strips were requested for investigation.The products have not been received at this time.If the products are returned in the future, a follow up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.
|
|
Manufacturer Narrative
|
The patient's and the clinic's test strips were not returned for investigation.As no products were returned, a specific root cause could not be determined.
|
|