• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WATCHMAN ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WATCHMAN ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL Back to Search Results
Model Number 10374
Device Problem Leak/Splash (1354)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 03/01/2023
Event Type  malfunction  
Event Description
It was reported there was a leak in the hemostatic valve.A left atrial appendage (laa) closure procedure was being performed.A watchman access system (was) double curve and a watchman laa closure device & delivery system (wds) were used.During the procedure the hemostatic valve could not be tightened enough.As a result of this there was blood leaking from the hemostatic valve while the devices were in the patient.The physician also noted that there was instability with operation of the pigtail catheter and had concerns of air introduction.The procedure was completed successfully with another was and the closure device was implanted in the laa of the patient.There were no complications to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN ACCESS SYSTEM
Type of Device
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key16556521
MDR Text Key311509750
Report Number2124215-2023-12266
Device Sequence Number1
Product Code NGV
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 03/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10374
Device Catalogue Number10374
Device Lot Number0028854547
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
Patient Weight65 KG
-
-