The device was received at olympus for evaluation.The evaluation results stated the error code was not confirmed by engineer.In addition, the following findings were also identified: due to corrosion on plug unit, the specified resolving power is not obtained.The image was abnormal, the image was shadowy due to scratched objective lens,the instrument channel tube port had leakage.The investigation is ongoing.Therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|
The customer reported to olympus that prior to the start of an unspecified surgical procedure, an error code e315 presented on the evis lucera elite colonovideoscope.It was confirmed that the patient was not under sedation.No death, injury or harm was reported.No procedural delay confirmed.Additional details have been requested regarding the reported event.At this time no additional information has been provided.
|