• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS CYFRA 21-1; TUMOR-ASSOCIATED ANTIGEN IMMUNOLOGICAL TEST SYSTEM.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS CYFRA 21-1; TUMOR-ASSOCIATED ANTIGEN IMMUNOLOGICAL TEST SYSTEM. Back to Search Results
Catalog Number ASKU
Device Problems Mechanical Problem (1384); High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2023
Event Type  malfunction  
Event Description
There was an allegation of questionable high results for multiple assays for one patient from cobas e 801 analytical unit serial number (b)(4).The results were reported outside of the laboratory and the doctor believed the results did not match the medical imaging results and questioned if there were interfering substances.Refer to the attachment to the medwatch for all patient data.Refer to the medwatches with the following a1 patient identifiers for all involved assays: (b)(6) for the ca 15-3 assay.(b)(6) for the cea assay.(b)(6) for the cyfra 21-1 assay.(b)(6) for the ca19-9 assay.(b)(6) for the ca 125 assay.(b)(6) for the afp assay.
 
Manufacturer Narrative
A sample from the patient was submitted for investigation.
 
Manufacturer Narrative
The patient sample was investigated and no cause for the result discrepancies could be found.The customer¿s results were confirmed.No product issue was found and the reagent performed within specifications.A general reagent-related issue could be excluded.
 
Manufacturer Narrative
Medwatch fields b3 and b6 were updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS CYFRA 21-1
Type of Device
TUMOR-ASSOCIATED ANTIGEN IMMUNOLOGICAL TEST SYSTEM.
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16567062
MDR Text Key311582701
Report Number1823260-2023-00863
Device Sequence Number1
Product Code OVK
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K160915
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age74 YR
Patient SexFemale
-
-