Catalog Number 07027559190
Device Problems High Test Results (2457); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2023
Event Type  malfunction  
Event Description
The initial reporter stated they received discrepant results for one patient sample tested with elecsys insulin on a cobas e 801 module (serial number (b)(4)).No questionable results were reported outside of the laboratory.The sample initially resulted in an insulin value of 37.2 uu/ml and it repeated as 19.2 uu/ml.
 
Manufacturer Narrative
The patient sample was repeated because the initial value was high and the patient had no result history.An additional repeat of the sample was performed and this value was similar to the repeat value of 19.2 uu/ml.Calibration and control recoveries were within expectations.The investigation could not identify a product problem.The cause of the event could not be determined.Medwatch field a3 has been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS INSULIN
Type of Device
IMMUNOREACTIVE INSULIN TEST SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16567198
MDR Text Key311825687
Report Number1823260-2023-00873
Device Sequence Number1
Product Code CFP
UDI-Device Identifier04015630940042
UDI-Public04015630940042
Combination Product (y/n)N
Reporter Country CodeJO
PMA/PMN Number
K001104
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07027559190
Device Lot Number68250901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexMale
-
-