ROCHE DIAGNOSTICS COAGUCHEK XS PT TEST PST; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 07671687019
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/20/2023
Event Type  malfunction  
Manufacturer Narrative
Reporter occupation is patient/consumer.The test strips were requested for investigation.The product has not been returned.If the product is returned in the future, a follow up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Product labeling states: "the coaguchek system uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas deviations can occur when compared to methods using other thromboplastins.However, those deviations between thromboplastins of different origin (e.G., rabbit based) are not specific to coaguchek products.Similar differences can be observed when a human recombinant thromboplastin based laboratory method is compared to other laboratory methods.".
 
Event Description
The customer complained of discrepant inr results with coaguchek inrange meter serial number unknown.The meter result was 2.6 inr.Customer repeated the test and the result was 1.9 inr.Customer repeated the test with the same finger used for the second test and the result was 3.1 inr.Customer's therapeutic range is 2.0-3.0 inr with a target of 2.5 inr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PT TEST PST
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16567894
MDR Text Key311825623
Report Number1823260-2023-00883
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K170960
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/31/2024
Device Catalogue Number07671687019
Device Lot Number62682413
Was Device Available for Evaluation? No
Date Manufacturer Received02/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexMale
-
-