MAKO SURGICAL CORP. UNKNOWN RESTORIS LATERAL INSERT - PKA 3.0 2.8MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
|
Back to Search Results |
|
Catalog Number UNK_OFL |
Device Problem
Naturally Worn (2988)
|
Patient Problem
Pain (1994)
|
Event Date 02/23/2023 |
Event Type
Injury
|
Event Description
|
Dr.Informed that he performed a mako lateral partial knee replacement on patient about 3.5 years ago.The patient was complaining of pain and he performed a scope yesterday on (b)(6) 2023 and noticed the tibial poly is "stained brown, it's pitting and when you look at it from the front you can see it's not flat and shows some concavity where the brown spot is".He also informed he found a chondral lesion on the patient's mfc.He stated that this implant has only been in for 3 1/2 years and this really shouldn't' happen.The scope he performed yesterday was due to the patient's continued pain.Case type / application: pka (mics) "right lateral".
|
|
Manufacturer Narrative
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted if additional information becomes available.Catalog numbers and lot codes of other devices listed in this report: size 1 lm/rl femur lot - unknown.Size 2 lm/rl tibial baseplate lot -unknown.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
|
|
Search Alerts/Recalls
|
|
|