Device evaluated by mfr.: synergy xd mr ous 3.00 x 38mm stent delivery system was returned for analysis.Examination of the crimped stent via scope found no issues with the stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of tip damage.A visual and tactile examination of the hypotube identified a break on the hypotube shaft located at 2.7 cm distal to the distal end of strain relief as well as multiple kinks.There was also a break identified in the hypotube bond at 36.7cm from the distal tip.The shaft polymer extrusion including the distal and mid-shaft sections were examined via scope and tactile examination.No issues were noted.No other issues were identified during the product analysis.
|
Reportable based on device analysis completed on 09mar2023.It was reported that difficulty opening/removing packaging material and shaft break occurred.A 3.00 x 38mm synergy xd drug-eluting stent was selected for use.However, when trying to remove the device from the packaging, it did not come out.Upon forcing and pulling it, the shaft broke at 1-2 cm from the hub.The procedure was completed successfully and no patient complications were reported.However, returned device analysis revealed a break on the hypotube shaft located at 2.7 cm distal to the distal end of strain relief.
|