• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REDAPT MODULAR SHELL 62MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REDAPT MODULAR SHELL 62MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Model Number 71352362
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problem Insufficient Information (4580)
Event Date 11/09/2020
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a patient complication identified through a review of clinical evidence from a post-market clinical follow-up activity that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event Description
On the pmcf titled dca-1000 accord ss clinical dataset, it was reported that, after undergoing left tha, one (1) patient presented a loose tha system.This adverse event was treated by performing a revision surgery combined with internal fixation on (b)(6) 2020, in which an accord cable system was implanted.A post-operative follow-up confirmed a good position without dislocation, wear nor evidence of failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDAPT MODULAR SHELL 62MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16583112
MDR Text Key311680960
Report Number1020279-2023-00601
Device Sequence Number1
Product Code MBL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182109
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Study,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71352362
Device Catalogue Number71352362
Device Lot Number18HM11733
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/16/2023
Date Device Manufactured08/15/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
40MM OX MDLR FM HD(PN:71342340,LN:20HM12189); R3 XLPE ACET LNR 40X62MM(PN:71338682,LN:15DM10585); RDPT 190MM RV STM SZ21(PN:71354471,LN:16MTM0103B); RDPT LCKNG SCRW 20MM(PN: 71354503,LN:UNK); RDPT LCKNG SCRW 25MM(PN:71354504,LN:19CM00105); RFLCTN SPH HD SCRW 30MM(PN:71332530,LN:20GB00006)
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age35 YR
Patient SexMale
-
-