Manufacturer's investigation conclusion: the reported event of an unknown alarm beeping from the controller while the display was not showing any alarms was not confirmed.The submitted controller event log file contained approximately 5 days of data (26feb2023 ¿ 03mar2023 per the timestamp).The log file did not capture any atypical alarms on 02mar2023; it was noted that the controller was connected to the mobile power unit (mpu) throughout the events captured on 02mar2023.The pump maintained a speed above the low speed limit throughout the log file.Additional information provided stated that the heartmate 3 system controller (serial number: (b)(6) ) was not exchanged and that the controller was not displaying low flow alarms when the reported unknown alarms occurred.Multiple requests for additional information were made to determine if the cause of the unknown alarm was determined, if any additional alarms activated, and if there were any issues observed with the patient cable on the mpu; however, no response was received.The root cause of the reported event could not be conclusively determined through this analysis.The device history records were reviewed and the records revealed that the heartmate 3 system controller, serial number (b)(6), was manufactured in accordance with manufacturing and qa specifications.The heartmate 3 system controller was shipped to the customer on 06apr2022.Heartmate 3 instructions for use (rev.C) section 7, entitled ¿alarms and troubleshooting¿, and heartmate 3 patient handbook (rev.D) section 5, entitled ¿alarms and troubleshooting¿ covers all alarms (visual and audible), and the actions to take if the alarm cannot be resolved.Heartmate 3 instructions for use (rev.C) section 8 entitled ¿equipment storage and care¿ and heartmate 3 patient handbook (rev.D) section 6 entitled ¿caring for the equipment¿ addresses how to properly care for and maintain the equipment for proper use.The heartmate 3 patient handbook (rev.D) cautions the user to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|