Manufacturer's investigation conclusion: the reported event of an unidentified alarm coming from the system controller was not confirmed.There was no photos or log files submitted for review.System controller, serial (b)(6), was not returned for analysis.The provided information indicated that there was an audible alarm coming from the controller without any information displaying on the controller.It is unknown how long the alarm was active for and therefore could have resolved before the patient looked at the displayed message on the controller.The battery clips were exchanged without resolving the issue.The alarm resolved after the controller was exchanged.A root cause for the reported event cannot be conclusively determined through this analysis.Device history records were reviewed and showed no deviations from manufacturing or quality assurance specifications.The patient handbook and instructions for use (ifu) also cautions the users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.Heartmate 3 ifu section 7-¿alarms and troubleshooting¿ and heartmate 3 patient handbook section 5-¿alarms and troubleshooting¿ explain how to troubleshoot all system controller alarms.No further information was provided.The manufacturer is closing the file on this event.
|
It was reported that on (b)(6) 2023, the patient heard an audible alarm with no information on the system controller.The battery clips were exchanged.On (b)(6) 2023, the patient experienced the problem again and heard an audible alarm with no information on the system controller.The system controller was exchanged which resolved the alarm.
|