Catalog Number 05168791190 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/23/2023 |
Event Type
malfunction
|
Event Description
|
There was an allegation of questionable gluc3 glucose hk gen.3 results for 1 patient urine sample on a cobas 8000 cobas c 701 module.The initial gluc3 results with and without auto-dilution were accompanied by a flag indicating the values were above the linearity/measuring range of the assay.This prompted the customer to perform a manual 1:5 dilution of the patient sample.The manual 1:5 dilution result was 18.45 g/l.The sample was sent to another laboratory and tested on another c701 analyzer.The sample was re-centrifuged and the result without dilution was 3.3 g/l.No questionable results were reported outside of the laboratory.The customer's analyzer serial number is (b)(4).
|
|
Manufacturer Narrative
|
The qc recovery data provided was acceptable.The investigation is ongoing.
|
|
Manufacturer Narrative
|
The calibration trace data provided was acceptable.The field service engineer found the analyzer was within specification.The investigation did not identify a product problem.The root cause of the event could not be determined.
|
|
Search Alerts/Recalls
|