Model Number CD2411-36Q
Device Problems Failure to Deliver Shock/Stimulation (1133); Incorrect Interpretation of Signal (1543)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/02/2023
Event Type  Injury  
Event Description
It was reported that the patient presented during clinical follow-up.Upon interrogation, it was noted that the implantable cardioverter defibrillator failed to provide high voltage output due to the device not classifying ventricular tachycardia with supraventricular tachycardia correctly.Programming changes were performed.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16595594
MDR Text Key311817195
Report Number2017865-2023-14490
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507615
UDI-Public05414734507615
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2021
Device Model NumberCD2411-36Q
Device Catalogue NumberCD2411-36Q
Device Lot NumberA000072222
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/17/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OPTISURE MRI; TENDRIL LEAD
Patient Outcome(s) Other;
Patient Age55 YR
Patient SexMale
-
-