Based on the investigation analysis, sensor could successfully link with a good transmitter but, the placement guide in the eversense app showed signal as "no signal." thus, the customer's complaint was reproduced.The transmitter along with the sensor was placed in 12 mm test fixture and sensor measurements were taken using sensor manager software.It was observed that the field current was (b)(4) adc units which is less than the threshold signal strength value of 412 adc units required to establish rf connection to the sensor.The root cause was determined to be low signal strength.As part of resolution, an rma was issued for sensor replacement.No further resolution was necessary for this complaint.H3.Device evaluated by manufacturer? yes.H6.Type of investigation updated to 10.H6.Investigation findings updated to 3237.H6.Investigation conclusions updated to 4307.
|