Model Number RONYX40026X |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Chest Pain (1776); Myocardial Infarction (1969); Obstruction/Occlusion (2422); Ischemic Heart Disease (2493); Thrombosis/Thrombus (4440); Insufficient Information (4580)
|
Event Date 01/28/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
During the index procedure three resolute onyx drug eluting stents were implanted in the right coronary artery (rca).Approximately 47 months post procedure the patient suffered from coronary atherosclerotic heart disease.The patient had paroxysmal chest pain for 6 years which got worse.The patient was admitted to hospital.It was stated that the event was a myocardial infarction involving the target lesion, the rca.Stent thrombosis was also confirmed by angiography in the three index procedure stents.The patient was on dual antiplatelet therapy within 24 hours prior to the event.The event was treated with a percutaneous coronary intervention.The patient recovered.
|
|
Manufacturer Narrative
|
Additional information: the cec adjudicated the revascularization as a clinically driven target lesion percutaneous coronary intervention in the rca.The cec adjudicated stent thrombosis as no event.The cec commented that there was limited information available but the patient did have occlusion of the rca which did not appear to have been successfully treated, thus occlusion predated this ae.Mi did not occur at time of event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|