Catalog Number 04625374160
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2023
Event Type  malfunction  
Manufacturer Narrative
Initial reporter occupation - occupation is patient/consumer.The test strips were requested for investigation.The product has not been returned.If the product is returned in the future, a follow up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Event Description
The customer complained of discrepant inr results with coaguchek vantus meter serial number (b)(4).The meter result was 3.7 inr.Customer tested again five minutes later and the result was 2.5 inr.The customer's therapeutic range is 2.5-2.7 inr.Customer tests on a weekly basis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16612245
MDR Text Key312297636
Report Number1823260-2023-00959
Device Sequence Number1
Product Code GJS
UDI-Device Identifier00365702127104
UDI-Public00365702127104
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K170960
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number04625374160
Device Lot Number62216021
Was Device Available for Evaluation? No
Date Manufacturer Received02/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CALCIUM; LEVOTHYROXINE; METOPROLOL; MULTIVITAMIN WITH IRON; OMEPRAZOLE; SIMVASTATIN; VITAMIN D; VITAMIN K; WARFARIN
Patient Age78 YR
Patient SexFemale
Patient Weight86 KG
-
-