Model Number A22040A
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2023
Event Type  malfunction  
Event Description
The end user facility contacted olympus to report that their inner shealth, for 26 fr.Outer sheath¿s sealing ring is damaged, the screw is lost, and the ceramic tip is damaged.The reported phenomena were found during reprocessing.No patient or user harm has been reported.No procedural or patient involvement have been reported.
 
Manufacturer Narrative
The subject device has been returned to an olympus repair center for evaluation and inspection, and the customer's reported issues have been confirmed.This report is providing information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over three years since the subject device was manufactured.Based on the results of the investigation, it is likely that general wear, tear, and mishandling led to the malfunction.A definitive root cause cannot be identified.This issue is addressed in the instructions for use (ifu): ¿inspecting the product visually inspect the product.Make sure that it has: no corrosion; no dents; no scratches; ceramic insulation at distal end; visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning: risk of injury.Impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product: if the product is damaged or does not function properly, contact an olympus representative or an authorized service center¿.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNER SHEATH, FOR 26 FR. OUTER SHEATH
Type of Device
RESECTOSCOPE SHEATH
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16616667
MDR Text Key312319032
Report Number9610773-2023-00877
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761029339
UDI-Public04042761029339
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K931994
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22040A
Device Catalogue NumberA22040A
Device Lot Number193W-0141
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2023
Was the Report Sent to FDA? No
Date Manufacturer Received02/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-