The end user facility contacted olympus to report that their inner shealth, for 26 fr.Outer sheath¿s sealing ring is damaged, the screw is lost, and the ceramic tip is damaged.The reported phenomena were found during reprocessing.No patient or user harm has been reported.No procedural or patient involvement have been reported.
|
The subject device has been returned to an olympus repair center for evaluation and inspection, and the customer's reported issues have been confirmed.This report is providing information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over three years since the subject device was manufactured.Based on the results of the investigation, it is likely that general wear, tear, and mishandling led to the malfunction.A definitive root cause cannot be identified.This issue is addressed in the instructions for use (ifu): ¿inspecting the product visually inspect the product.Make sure that it has: no corrosion; no dents; no scratches; ceramic insulation at distal end; visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning: risk of injury.Impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product: if the product is damaged or does not function properly, contact an olympus representative or an authorized service center¿.Olympus will continue to monitor the field performance of this device.
|