• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AGILE ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AGILE ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL Back to Search Results
Model Number M00517460
Device Problems Break (1069); Positioning Failure (1158); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/13/2023
Event Type  malfunction  
Manufacturer Narrative
Impact code f1001 is being used to capture the reportable issue of aborted/cancelled procedure.
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2023 that an agile esophageal fully covered stent was used to treat an esophageal cancer during an esophagogastroduodenoscopy (egd) with stent placement procedure performed on (b)(6) 2023.The patient's anatomy was tight and was not dilated prior to stent placement procedure.During the procedure, when the stent was being deployed, it was noticed that the outer blue sheath was kinked.Subsequently, a snap was heard and the stent was unable to be fully deployed.The stent was reconstrained and was fully covered by the outer sheath when it was removed from the patient.The procedure was not completed due to device availability.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGILE ESOPHAGEAL
Type of Device
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key16620454
MDR Text Key312101952
Report Number3005099803-2023-01609
Device Sequence Number1
Product Code ESW
UDI-Device Identifier08714729973089
UDI-Public08714729973089
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K180144
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00517460
Device Catalogue Number1746
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-