MEDTRONIC NEUROMODULATION IMPLANTABLE NEUROSTIMULATOR; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number NEU_INS_STIMULATOR |
Device Problems
Malposition of Device (2616); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Dysphagia/ Odynophagia (1815); Intracranial Hemorrhage (1891); Unspecified Infection (1930); Nausea (1970); Pain (1994); Vomiting (2144); Insufficient Information (4580)
|
Event Date 11/28/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
This value is the average age of the patients reported in the article as specific patients could not be identified.This value reflects the gender of the majority of the patients reported in the article as specific patients could not be identified.Please note that this date is based off of the date of publication of the article as the event dates were not provided in the published literature.It was not possible to ascertain specific device information from the article or to match the events reported with previously reported events.Correspondence will be sent to the author of the article inquiring about individual patient information and additional information regarding the reported events.Concomitant medical products: other relevant device(s) are: product id: neu_unknown_lead, serial/lot #: unknown.Product id: neu_unknown_lead, serial/lot #: unknown.Product id: neu_unknown_lead, serial/lot #: unknown.Product id: neu_unknown_lead, serial/lot #: unknown.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Powers, andrew y.Md; nguyen, michael ba; phillips, katharine ba; mackel, charles e.Md, jd; alterman, ron l.Md.Complications related to deep brain stimulation lead implantation: a single-surgeon case series.Operative neurosurgery 24(3):p 276-282, march 2023.| doi: 10.1227/ons.0000000000000513.Abstract: background: deep brain stimulation (dbs) is the mainstay of surgical treatment for movement disorders, yet previous studies have shown widely varying complication rates.Given the elective nature of dbs surgery, minimizing surgical complications is imperative.Objective: to evaluate short-term and long-term complications related to dbs lead implantation surgeries performed by an experienced surgeon and provide an updated benchmark comparison for other dbs centers and alternative therapies.Methods: a retrospective chart review of patients who underwent dbs lead implantation surgery by a single surgeon at our institution between 2012 and 2020 was conducted.Demographic and clinical data including surgical complications were collected.A kaplan-meier survival analysis was used to evaluate the cumulative risk of lead revision or removal over time.Associations between patient characteristics and various complications were evaluated.Results: four hundred fifty-one dbs leads were placed in 255 patients.Thirteen leads and 11 patients required revision.In total, 3.6% (95% ci [1.3%-5.9%]) of patients required revision at 1 year and 4.8% (95% ci [1.9%-7.6%]) at 5 years, with per-lead revision rates of 2.3% (95% ci [0.9%-3.6%]) and 3.3% (95% ci [1.5%-5.1%]), respectively.Less common diagnoses such as tourette syndrome, post-traumatic tremor, and cluster headache trended toward association with lead revision or removal.Conclusion: dbs performed by an experienced surgeon is associated with extremely low complication rates.Reported events: 1.Eight patients experienced lead removal or revision due to either infection or malposition.Some of the infections occurred within a 7-month period in 2019.This clustering suggest that unknown exogenous factors may have contributed to these infections as the provider's surgical technique remained quite consistent throughout the study period. 2.One patient experienced prolonged hospitalization due to intracranial hemorrhage resulting in transient motor deficits that resolved over time without intervention. 3.Two patients experienced prolonged hospitalization due to one of the following complications: pain, dysphagia, or nausea/emesis. 4.Five patients' surgery was aborted due to stimulation side effects. the higher rate of aborted surgeries may be due to a more con servative approach when stimulation effects suggest suboptimal lead placement that cannot be corrected with appropriate adjustments in lead position.
|
|
Search Alerts/Recalls
|
|
|