MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA PLUS VASCULAR COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA PLUS VASCULAR COVERED STENT Back to Search Results
Catalog Number AASLE10080
Device Problems Loss of or Failure to Bond (1068); Positioning Failure (1158)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2023
Event Type  malfunction  
Manufacturer Narrative
A voluntary recall has been initiated for the covera vascular covered stent which was product catalog/lot number specific.Reportedly the covera vascular covered stent has the potential to exhibit deployment issues (i.E., failure to deploy the covered stent) due to slide block bond failures in the device handle.Full or partial failure of the product to deploy is most likely to require percutaneous replacement with existing access.In rare cases, iatrogenic vascular injury may occur as the result of device withdrawal, manipulation, or abrupt/unexpected movements, particularly in the context of difficult or partial stent deployment or mispositioning.Adjunctive endovascular maneuvers may be indicated.There have been no cases where an open surgical procedure has been required.To date, there have been no reported patient injuries associated with deployment issues.As a result of the field action, this event is being reported as a malfunction reportable event.The catalog number identified in catalog# has not been cleared in the us but is similar to the covera plus vascular covered stent system that are cleared in the us.The pro code and 510 k number for the covera plus vascular covered stent system are identified in common device name and catalog#.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiration date: 08/2023).
 
Event Description
It was reported that during a stent graft placement procedure via femoral access, the large thumbwheel allegedly failed to function after unlocking safety button and the stent allegedly failed to deploy.There was no reported patient injury.
 
Manufacturer Narrative
H10: a voluntary recall has been initiated for the covera vascular covered stent which was product catalog/lot number specific.Reportedly the covera vascular covered stent has the potential to exhibit deployment issues (i.E., failure to deploy the covered stent) due to slide block bond failures in the device handle.Full or partial failure of the product to deploy is most likely to require percutaneous replacement with existing access.In rare cases, iatrogenic vascular injury may occur as the result of device withdrawal, manipulation, or abrupt/unexpected movements, particularly in the context of difficult or partial stent deployment or mispositioning.Adjunctive endovascular maneuvers may be indicated.There have been no cases where an open surgical procedure has been required.To date, there have been no reported patient injuries associated with deployment issues.As a result of the field action, this event is being reported as a malfunction reportable event.H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the covera plus vascular covered stent system that are cleared in the us.The pro code and 510 k number for the covera plus vascular covered stent system are identified in d2 and g4.H10: manufacturing review: the lot history records were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the stent delivery system of a covered stent was returned for evaluation.The slide block, a force transmitting component was no longer connected to the proximal sheath.The disconnection of the slide block is considered to have led to the reported impossibility to deploy the stent.Based on evaluation of the sample the inability of an adhesive joint (slide block/tether/diving sheath) to withstand tension force during deployment is confirmed.A definitive root cause could not be determined based upon the available information.The intended use of the device in the portal vein represents off-label use.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.Based on the instructions for use supplied with this product delivery system specific events that could be associated with clinical complications include but are not limited to failure to deploy and high deployment forces.Regarding correct deployment the instructions for use states '(¿) maintain a stationary hold on the white stability sheath during covered stent deployment (¿).Hold the white stability sheath as close as possible to the introducer without touching the dark brown moving catheter of the distal catheter assembly.Maintain the remainder of the white stability sheath (.) relaxed and avoid tension.' regarding preparation the instructions for use states: 'pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated.' the instructions for use further state: 'the covera plus vascular covered stent is indicated for the treatment of stenoses in the upper extremity venous outflow of patients dialyzing with an arterio-venous (av) access graft or fistula and for the treatment of atherosclerotic lesions in iliac and femoral arteries with a reference vessel diameter of 4.5 mm to 9 mm'.The intended use of the device in the portal vein represents off-label use.H10: d4 (expiration date: 08/2023), g3, h6 (device).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event Description
It was reported that during a stent graft placement procedure via femoral access, the large thumbwheel allegedly failed to function after unlocking safety button and the stent allegedly failed to deploy.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COVERA PLUS VASCULAR COVERED STENT
Type of Device
VASCULAR COVERED STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16626076
MDR Text Key312430278
Report Number9681442-2023-00104
Device Sequence Number1
Product Code PFV
UDI-Device Identifier00801741136283
UDI-Public(01)00801741136283
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
P170042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 04/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAASLE10080
Device Lot NumberANFV0310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/14/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberRES 91359
Patient Sequence Number1
-
-