Catalog Number 05.001.202
Device Problems Leak/Splash (1354); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2023
Event Type  malfunction  
Event Description
It was reported by germany that during service and evaluation, it was determined that the power module device had a liquid leak.It was further observed that the device had foreign substance/debris/cleaning/sterilization and fluid ingress.It was further determined that the device failed pretest for general condition and lever function, check for external cleanness and foils of liquid damage, information button and self-test and check liquid indicator.It was noted in the service order that the device had an undetermined malfunction.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese, or its employees that the report constitutes an admission that the product, depuy synthese, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device leaking liquid during service and evaluation was confirmed.The assignable root cause was determined to be traced to user, which is user error.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER MODULE FOR TRAUMA RECON SYSTEM
Type of Device
BATTERY, REPLACEMENT, RECHARGEABLE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key16628530
MDR Text Key312294992
Report Number8030965-2023-03749
Device Sequence Number1
Product Code MOQ
UDI-Device Identifier07611819977822
UDI-Public07611819977822
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.202
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2023
Date Manufacturer Received03/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/05/2016
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-