• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUROPACE, INC. NEUROPACE RNS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number DL-330-10-K
Device Problems Signal Artifact/Noise (1036); Impedance Problem (2950)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2022
Event Type  malfunction  
Event Description
During a routine review of the patient ecog data, high impedances and signal artifact were observed on lead 1.These unusual signals were indicative of a lead break.On (b)(6) 2023, the treating physician turned off detection and stimulation for that lead.On (b)(6) 2023, the patient underwent a system explant (the neurostimulator and four leads, including the suspect lead).On (b)(6) 2023, the patient was reimplanted with the rns and four leads.The rns system was explanted and replaced to reposition the leads and improve effectivity of the rns system.
 
Manufacturer Narrative
(b)(4).Neuropace is pending return of the explanted product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502796389
MDR Report Key16628740
MDR Text Key312433057
Report Number3004426659-2023-00014
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005045
UDI-Public010085554700504517211015
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDL-330-10-K
Device Catalogue Number1007605
Device Lot Number26785-1-1-1
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age24 YR
Patient SexMale
-
-