BOSTON SCIENTIFIC CORPORATION AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM; PANCREATIC STENT, COVERED, METALLIC, REMOVABLE
|
Back to Search Results |
|
Device Problems
Premature Activation (1484); Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/17/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The complainant was unable to provide the suspect device upn and lot number; therefore, the lot expiration and device manufacture dates are unknown.Imdrf device code a150103 captures the reportable event of first flange of the axios stent prematurely deployed.
|
|
Event Description
|
It was reported to boston scientific corporation on (b)(6), 2023, that an axios stent and electrocautery enhanced delivery system was to be implanted during a procedure performed on (b)(6) 2023.The issue has been occurring at irregular intervals since january.During the procedure, locking the stent deployment hub was often not smooth and the physician had to wiggle the deployment hub to lock it.Subsequently, deployment of the first flange of the axios stent was abrupt and not predictable.No patient complication was reported due to this event.
|
|
Search Alerts/Recalls
|
|
|