• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 PPS LTD ACET SHELL 50D; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 PPS LTD ACET SHELL 50D; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Failure of Implant (1924); Pain (1994); Osteolysis (2377); Ambulation Difficulties (2544); Foreign Body In Patient (2687)
Event Date 03/08/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: cat# 650-1057 lot# 2958832 cer bioloxd option hd 36mm.Cat# 650-1065 lot# 2959085 cer option type 1 tpr sleve -3.Cat# 00625006530 lot# 64403729 bone scr 6.5x30 self-tap.Cat# 010000856 lot# 6443904 g7 neutral e1 liner 36mm d.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported the patient underwent an initial left total hip arthroplasty that was subsequently revised approximately four (4) years later due to pain, noise, instability, osteolysis, and loosening.During the revision while placing the cage noted a screw broke and was unable to retrieve broken part.The stem was well fixed and remained implanted.The shell, liner, head, and taper sleeve were exchanged without complications.Attempts have been made and no further information has been provided.
 
Event Description
No further information at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: x-ray found acetabular loosening superomedial breaching medial wall, ct ¿ osteolysis, acetabular protusio consistent with loosening, nondisplaced fracture of inferior medial wall, acetabular component malposition.Pain in left hip for 15 months along with clicking, instability, snapping/popping, no history of falls in last 12 months, ambulating with assistive device.Noted acetabular bone loss superior and medial.No complications, noted procedure 100% took more time due to large acetabular bone defect, need for cage and additional fixation.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 PPS LTD ACET SHELL 50D
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16643144
MDR Text Key312299198
Report Number0001825034-2023-00705
Device Sequence Number1
Product Code PBI
UDI-Device Identifier00880304524217
UDI-Public(01)00880304524217(17)290522(10)6549335
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number010000662
Device Lot Number6549335
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexMale
Patient Weight81 KG
-
-