• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH SCRDRIVER SHAFT T8 SELF-HOLDING; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH SCRDRIVER SHAFT T8 SELF-HOLDING; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL Back to Search Results
Catalog Number 03.617.902
Device Problem No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: additional product code: kwq part number: 03.617.902 lot number: 400p269 manufacturing site: haegendorf release to warehouse date: 04.10.2021 a manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the scrdriver shaft t8 self-holding was worn from the distal tip.The observed condition was consistent as an end of life indicator for the device.No other issues were identified.After a visual inspection per guidance provided in windchill document, it was determined that the reusable instrument device was worn from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the scrdriver shaft t8 self-holding would have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: the following document was reviewed: windchill document.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in colombia as follows: it was reported on an unknown date that the device was received as a blind unit from colombia on january 19th.There is no allegation reported against the device.This report is for one (1) scrdriver shaft t8 self-holding this is report 1 of 1 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCRDRIVER SHAFT T8 SELF-HOLDING
Type of Device
INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK HÄGENDORF (CH)
im bifang 6
haegendorf 4614
SZ   4614
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16644407
MDR Text Key312333612
Report Number8030965-2023-03866
Device Sequence Number1
Product Code OVE
UDI-Device Identifier07611819969742
UDI-Public(01)07611819969742
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K152239
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number03.617.902
Device Lot Number400P269
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2023
Date Manufacturer Received03/28/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
SCREW Ø2.7 L14 F/CERVIC-DISTRACTOR SST; SCREW Ø2.7 L16 F/CERVIC-DISTRACTOR SST
-
-