The mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - evaluation found no device deficiencies that would have contributed to the reported complaint and was unable to duplicate the reported issue.The patient's skull was fractured under 40lbs of pressure, and the unit has slight movement in the lock, but when properly positioned and put under pressure it will not move.The unit passes all specific functional testing.The unit has no service history, so it is recommended that a preventative maintenance be performed at this time.Unrelated to the reported issue, investigation showed that the lock had both rotational and lateral movement and a residue buildup was present.Upon disassembly, it was noted that the index knob and the lock needed new components added to replace worn internal parts.As a result, the unit was machined to have heli-coils added to large starburst threads; new components were added to replace worn internal parts, and general maintenance and cleaning were performed.Root cause - the reported complaint is not confirmed.Probable root cause is improper placement or use of the skull clamp on the patient.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|