During a procedure, there were intermittent ep-4 is off or not connected messages resulting in cancellation of the procedure.The ep-4 stimulator was swapped out.While pacing at an interval of 250ms, the pacing could not be halted with the touchscreen or the workmate claris computer.The ep-4 stimulator was power cycled but when powered back on, it resumed pacing.The channels were deselected on the workmate claris but pacing spikes were still displaying.The channel switches on the ep-4 stimulator were turned off but the issue still did not resolve.The entire system was also power cycled but the issue did not resolve.The procedure was then abandoned.After the procedure during troubleshooting replacing the partially broken fiber optic cable causing intermittent connection dropout appears to have resolved the issue.
|