ST PAUL LEVEL 1 HOTLINE LOW FLOW SYSTEMS; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-90
Device Problems Electrical Power Problem (2925); Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A product sample was received for evaluation.Visual and functional testing were performed.Physical observation showed broken plat clip, cracked enclosure, tank cover, outdated printed circuit board (pcb), and power switch.Started with a visual inspection then filled tank with water, attached temperature check, plugged in line cord, and turned on the power switch.The device failed the electrical test.The cause was slice in the power cord.The root cause of the reported issue was customer use.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.Udi information are unknown.
 
Event Description
It was reported that the device is tripping circuit breakers during pre-test.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE LOW FLOW SYSTEMS
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16660779
MDR Text Key312530210
Report Number3012307300-2023-03359
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-