• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® PROPATEN® VASCULAR GRAFT; PROSTHESIS, VASCULAR GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® PROPATEN® VASCULAR GRAFT; PROSTHESIS, VASCULAR GRAFT Back to Search Results
Catalog Number HT076080
Device Problem Complete Blockage (1094)
Patient Problem Obstruction/Occlusion (2422)
Event Date 10/23/2015
Event Type  Injury  
Manufacturer Narrative
No patient specific details have been provided.Therefore, the patient initials reflect the w.L.Gore internal case number.Heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.Code b13: further information related to the procedure, the incident and the patient condition was requested from the hospital and was captured in this report.Code b20: the device remains implanted in the patient.Therefore a device evaluation could not be performed.Review of the manufacturing records is currently ongoing and the result will be provided in the final report.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Reportedly on (b)(6) 2015, this patient underwent a bypass surgical treatment for an occlusive peripheral arterial disease in which a gore propaten vascular graft was used.Reportedly, on (b)(6) 2015, an adverse event termed "thrombosis of the gore bypass" was discovered.A reintervention was performed and the adverse event resolved on the same date via a reintervention.The primary relationship was confirmed as disease related.
 
Manufacturer Narrative
H6: corrected imdrf codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® PROPATEN® VASCULAR GRAFT
Type of Device
PROSTHESIS, VASCULAR GRAFT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WEST B/P
1505 n. fourth street
flagstaff AZ 86004
Manufacturer Contact
wilson okeke
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key16661861
MDR Text Key312498339
Report Number2017233-2023-03837
Device Sequence Number1
Product Code DSY
Combination Product (y/n)Y
Reporter Country CodeFR
PMA/PMN Number
K062161
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/03/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/13/2019
Device Catalogue NumberHT076080
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/14/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age66 YR
Patient SexMale
-
-