H10: through follow-up communication livanova learned that the water leak was internal to the device.Indeed, a livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue: water was leaking from the t-connection on the internal blue tubing.In oder to solve the issue, blue tubing set was replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Therefore, being the water leak internal to the device, it is unlikely to cause serious injury to the user.Thus, the event has been re-assessed as not reportable.
|